DEVELOPMENT AND VALIDATION OF A STABILITY INDICATING RP-HPLC METHOD FOR SIMULTANEOUS DETERMINATION OF VALSARTAN AND SACUBITRIL IN TABLET DOSAGE FORM.
M. Madan Mohan Reddy*, D. Gowri Sankar and JVLN Seshagiri Rao
ABSTRACT
Objective: A New method was established for simultaneous
estimation of Sacubitril And Valsartan by RP-HPLC method.
Methods: Chromatographic separations were carried using Spurcil
C18 (4.6 x 250mm, 5m) (Dikma column) column with a mobile
phase composition of 0.1% OPA buffer and Acetonitrile(50:50) have
been delivered at a flow rate of 1ml/min and the detection was carried
out using waters HPLC auto sampler, separation module 2695 with
PDA detector at wavelength 237 nm. Results: The retention time for
Sacubitril and Valsartan were 3.119 and 6.851 minute respectively.
The correlation coefficient values in linearity were found to be 0.999
and concentration range 12-60 μg/ml for Sacubitril and 13-65μg/ml for
Valsartan respectively. For accuracy the total recovery was found to be
99.98% and 100.28% for Sacubitril and Valsartan. The LOD and LOQ
concentration for Sacubitril were found to be 0.02 μg/ml and 0.05 μg/ml and LOD and LOQ
for Valsartan were found to be 0.01μg/ml and 0.04 μg/ml. The force degradation studies were
performed and the results are within the limits. Conclusion: The results of study showed that
the proposed RP‐HPLC method is a simple, accurate, precise, rugged, robust, fast and
reproducible, which may be useful for the routine estimation of Sacubitril and Valsartan in
pharmaceutical dosage form.
Keywords: Sacubitril, Valsartan, RP?HPLC, Validation.