UV SPECTROPHOTOMETRIC ESTIMATION OF ACYCLOVIR IN BULK AND TABLET DOSAGE FORM USING AREA UNDER CURVE METHOD
Bhagyashri P.Pawaskar*, SugandhaV. Mulgund
ABSTRACT
This paper describes development and validation of UV
spectrophotometric method for the estimation of Acyclovir in tablet
formulation using area under curve method. The standard and sample
solutions of Acyclovir were prepared in distilled water. The area under
curve in between 241to 251 nm was measured. The method followed
linearity in the range of 2-12μg/ml with correlation coefficient value of
0.999. This method was validated for various parameters according to
ICH guideline Q2 (R1). Satisfactory values of Percent relative
standard deviation for the intra-day and inter-day precision indicated
that method was precise. The mean percentage recovery of analyte
was found to be 99.41%. LOD and LOQ were calculated as 0.3134μg/ml and 0.9515μg/ml,
respectively.
Keywords: Acyclovir, UV spectrophotometry, Area under curve, Validation.