METHOD DEVELOPMENT AND VALIDATION OF METHOTREXATE DRUG BY RP-HPLC
A Santosh kumar Sreevatsav*, N. Shiva Prasad, D.Akhilesh Kumar,
Sanjay Kumar M, M.Satish Kumar
ABSTRACT
A simple and reproducible method was developed for Methotrexate by
Reverse phase high performance liquid chromatography (RP-HPLC).
Methotrexate was separated on C-18 column (1000mm x 6mm x 5μ),
using potassium di hydrogen orthophosphate at the UV detection of
303nm. Isocratic elution of Acetonitrile and buffer as mobile phase
with various ratios and flow rates. Eventually 92:8 v/v buffer and
acetonitrile was being set with the flow rate of 1.4 ml/min. The
statistical validation parameter such as selectivity (or) specificity,
linearity, precision,accuracy, robustness were checked. The recovery
and assay studies of method were within 97% to 102%. Indicates that
the proposed method can be adoptable for Quality control analysis of method.
Keywords: Methotrexate, RP-HPLC, Validation, acetonitrile
[Download Article]
[Download Certifiate]